Why is there insufficient gas flow in my NEW BRUNSWICK SCIENTIFIC Laboratory Equipment?
C
Christopher AlvarezAug 14, 2025
Insufficient gas flow can occur due to several reasons. The inlet or exhaust sterile air filter might be wet or clogged, in which case you should replace it. Ensure that the air pressure is within the specified range. Verify that the control mode for DO and pH is set to AUTO and that the GasFlo loop is ON. Also, confirm that the Air loop is in O2 Enrichment mode and that the DO cascades are Enabled.
E
ejordanAug 17, 2025
What to do if my NEW BRUNSWICK SCIENTIFIC BioFlo CelliGen 115 Laboratory Equipment is leaking water?
M
Manuel BrooksAug 17, 2025
If your unit is leaking water, the inlet water pressure may be too high. Try lowering the pressure to stay within the recommended range. Also, check all inlet hose connections to ensure they are tight; tighten them if necessary.
J
joel79Aug 20, 2025
What to do if the touchscreen is not responding on my NEW BRUNSWICK SCIENTIFIC Laboratory Equipment?
D
David HebertAug 21, 2025
If the touchscreen on your equipment is not responding, try calibrating the touchscreen.
C
catherine50Aug 25, 2025
What to do if the agitator does not turn in my NEW BRUNSWICK SCIENTIFIC Laboratory Equipment?
L
Linda RogersAug 25, 2025
If the agitator does not turn, or turns slowly, the motor drive coupling may not be installed properly. Refer to the motor adaptation instructions in this manual, and then check the coupling. You may need to remove or replace the O-ring. Also, ensure the motor is plugged into the cabinet receptacle, remembering to turn off the main power before connecting the plug.
L
laurensantosAug 27, 2025
Why won't my NEW BRUNSWICK SCIENTIFIC BioFlo CelliGen 115 heat up?
D
Darlene DominguezAug 27, 2025
If the unit will not heat up, first make sure it was primed at start-up. Ensure the temperature probe is plugged into the vessel thermowell. Check that the water pressure is within the recommended range. Verify the correct connection (click to lock) of the water inlet and outlet lines on the vessel heat exchanger. Finally, hit the reset button on the hot plate, if there is one.
L
ljacobsAug 30, 2025
What to do if pH probe readings are erratic on my NEW BRUNSWICK SCIENTIFIC Laboratory Equipment?
G
gomezjenniferAug 31, 2025
If you're getting erratic pH probe readings, recalibrate the probe, carefully following instructions in this manual. Check for a secure connection. If you have a gel-filled probe, it may need replacement. If it's liquid-filled, it may need a refill of electrolyte. The probe cable may also need replacement.
A
Aaron SmithSep 3, 2025
What to do if DO probe readings are erratic on my NEW BRUNSWICK SCIENTIFIC BioFlo CelliGen 115?
L
Lauren StoneSep 3, 2025
If the DO probe readings are erratic, recalibrate the probe, carefully following the instructions in this manual. You might also need to recharge the probe, following the manual's instructions. The probe may need a new membrane and a refill of electrolyte. Check for a secure connection. As a last resort, replace the probe cable or DO probe.
F
fadamsSep 6, 2025
What to do if my NEW BRUNSWICK SCIENTIFIC BioFlo CelliGen 115 probe does not hold calibration?
S
Steven PayneSep 6, 2025
If the probe does not hold calibration, the probe may be defective and need replacement. The pH/DO board may also be defective, in which case you should call for service.
T
Tami MartinSep 10, 2025
How to fix a negative readout value on NEW BRUNSWICK SCIENTIFIC Laboratory Equipment?
J
Jason CrawfordSep 10, 2025
If the readout shows a negative value (typically –225° C), inspect the temperature probe for any obvious damage and replace it if necessary. Also, ensure that the temperature probe is correctly connected to the cabinet jack.