2.4
AIR
1
OXYGEN
MIXER
The air/oxygen mixer has the dual function of mixing air and oxygen to
a desired
F102 and regulating the output mixture to a preset pressure.
The pressure of the air and oxygen supply gases are lowered to 2.0
kg/cm2 (28.5 psi) by means of a diaphragm and regulator assembly.
The regulated gases are then mixed and supplied at the desired
F102.
The input pressure of the gases must be between 40
-
60 psi for the
mixer to maintain accurate
F102 and regulate the output pressure.
Figure
2.1
displays the four output ports of the mixer.
(1)
Main Flow port:
Supplies the main flow of the unit.
(2) Continuous Flow port:
Supplies bleed flow to the atmosphere. This flow is the
minimum flow required out of the mixer to maintain the
F102
within specifications.
(3)
Auxiliary Flow port:
Supplies the Spontaneous flow and Auxiliary flow to the
unit.
(4) Air Port:
Supplies gas to the pneumatic control circuit and is equal to