11. Compare result to Control Range printed
on Strip vial label for Control Level you
are using.
If result is in range, System can be used
for testing blood.
If result does not fall within range, repeat
test using a new Strip.
Note: Control Test result shows
the Control Symbol in the Display.
Note:
When using the meter for the first time,
check the unit of measure in the meter
Display aer performing a control
or blood test. Result must display in
the correct unit of measure (mg/dL).
If the result does not display in the
correct unit of measure, DO NOT use
the meter and DO NOT use the meter
result to determine treatment.
Caution! If Control Test result is outside range, test again.
If result is still outside range, System should not be
used for testing blood. Call for assistance (see cover
for phone number).
12. Aer result is shown, Strip Release Button
ashes. Hold Meter with Test Strip pointing
down. Press Strip Release Button to release
and discard Strip in appropriate container.
Meter turns o.
Note: Removing Strip before result
displays cancels the test. An error message appears and
the result is not stored in Memory. Retest with a new
Strip and do not remove before result is displayed.
20
Control Symbol
Test Strip
Vial Label
Glucose Control
Bottle Label