Nipro TrueResult - Error Messages and Displays: Codes E1-E5; E-1 Temperature Error; E-2 Sample or Strip Error; E-3 Strip or Sample Placement Error

Nipro TrueResult
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Messages
Display
Reason
Temperature
Error
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Strip,
Strip outside of
vial too long,
Sample on top
of Strip.
Meter
Error
Test Strip
Error, Very
high blood
glucose result
(higher than
600 mg/dL)
Action
Move Meter and Strips
to area between
50-104ºF; wait 10 minutes
for System to reach
room temperature before
testing.
Retest with new
TRUEtest Test Strip and
larger sample.
Repeat with new Strip.
Make sure sample is
touched to edge of
Strip (not top). If error
persists, call for assistance.
Call for
assistance.
Retest with new Strip. If error
persists, call for assistance.
If you have symptoms such
as fatigue, excess urination,
thirst, or blurry vision, follow
your healthcare professional's
advice for high blood glucose.