IMPORTANT INFORMATION:
For the most accurate results using TRUEresult:
Read all instructions before testing.
TRUEresult is an in vitro (outside body)
quantitative system that is used for self-testing and
point-of-care testing of human whole blood only.
Meter displays results as plasma values.
Use only TRUEtest Test Strips and TRUEtest
Glucose Control with TRUEresult Meter.
Do not use for diagnosis of diabetes or for
testing blood glucose in newborns.
Perform Control Tests before performing a blood
glucose test for the first time (see Getting Started).
Note: ree levels of TRUEtest Glucose Control
Solution are available for Quality Control
Testing. It is important to perform Control
Tests with more than one level of Control to
assure your System is working properly and
your testing technique is good. Contact place
of purchase or call for assistance to obtain
Control.
Remove only one Strip at a time from vial when
testing. Recap vial immediately aer removing
Strip.
When using the meter for the first time, check
the unit of measure in the meter Display aer
performing a control or blood test. Result must
display in the correct unit of measure (mg/dL).
If the result does not display in the correct unit of
measure, DO NOT use the meter and DO NOT
use the meter result to determine treatment.