THIS GUIDE IS FOR REFERENCE ONLY. Refer to the Operator’s Manual for complete
information, warnings, cautions and specifications for the product you are using.
Use ONLY with compatible Nonin-branded pulse oximeter sensors.
P/N 9484-001-01
Model 7500/7500FO SpO
2
Pulse Oximeters
DEFAULT ALARM LIMITS and TIME/DATE
Default Alarm Limits are preset on the Model 7500 oximeters, which may be
adjusted prior to use. Default limits are:
High SpO
2
– Off High Pulse Rate – 200 BPM
Low SpO
2
– 85% Low Pulse Rate – 50 BPM
Caution: Always verify alarm settings when turning on the unit.
To set Time, Date and Alarm Limits in standard mode, turn the oximeter on
and follow these steps:
Press the Limits Button.
Use the + and – buttons to move the value up or down.
Press Limits again to set the value and move to the next setting.
You can also use the Limits button to review current settings by pressing
through each value. You will then see and reset that value with each press.
Patient Security Mode locks Limits, Time, Date and other values. See the
Operator’s Manual on Setting and Releasing Patient Security Mode if in use.
1. Select the proper size Nonin-branded sensor. Place sensor on patient
prior to connecting to the oximeter.
2. Connect Model 7500 Oximeter to wall power, or check battery
level prior to use. Recharge if required. Oximeter runs at least
16 hours on a full battery.
3. Insert the sensor connector into the
sensor port on the oximeter.
4. Turn the oximeter on with the On/Off /
Standby button. The oximeter will
go through a self-check, then begin
displaying oxygen saturation (
SpO
2
) and
pulse rate.
Do not use oximeter if any LED segments
are missing or displays do not appear
as expected. Refer to the Operator’s Manual
for instructions.
Alarms are disabled for two minutes
to allow time for sensor placement and adjustment.
Model 7500FO for MRI environments requires special sensors and
precautions. Refer to the Model 7500FO Operator’s Manual.
Model 7500/ 7500FO
SpO
2
Pulse Oximeters
SpO
2
3
4
Pulse
Rate