Operators Manual
LifeSense
®
LS1-9R
Vital Signs Monitor
Capnography/Pulse Oximeter
English

Table of Contents

Questions and Answers

  • K
    Kevin MccoyAug 10, 2025
    How to fix low EtCO alarm 2 on Nonin Medical Equipment?
    • J
      jacob51Aug 10, 2025
      If the Nonin device shows a low EtCO alarm 2, even when the patient's EtCO is suspected to be normal, start by checking the patient's status. Inspect the moisture trap and filter, and replace them if necessary. Also, check the sample line connector and visually inspect the sample line for any signs of damage. Finally, check the Nafion tubing connection. It is possible that an air leakage has occurred in the sample line, Nafion tubing, moisture trap, or internally.
  • A
    Alyssa FerrellAug 25, 2025
    What to do if Nonin Medical Equipment shows WARM UP with alarms?
    • K
      Kimberly MendozaAug 25, 2025
      If the Nonin device displays 'WARM UP' with alarms, first ensure the filter is correctly in place and replace it if needed. Next, perform a calibration and gas verification to confirm the device's performance. An equipment fault may be suspected if readings are out of range.

Nonin LifeSense LS1-9R Specifications

General IconGeneral
CategoryMedical Equipment
ModelLS1-9R
BrandNonin
Product LineLifeSense
Measurement ParametersSpO2, Pulse Rate
Operating Temperature5°C to 40°C
Device TypePulse Oximeter
SpO2 Range0% to 100%
Pulse Rate Range18 to 300 BPM
AccuracySpO2: ±2 digits, Pulse Rate: ±3 digits
Battery TypeAAA
Dimensions64mm x 34mm x 28mm
Storage Temperature-20°C to 60°C
Humidity Range10% to 95% (non-condensing)