Operator’s Manual
Model 3150
WristOx
2
®
Pulse Oximeter
BLE and USB
0123
English

Table of Contents

Questions and Answers

  • A
    Anna ShafferSep 23, 2025
    Why are %SpO and heart not displaying on Nonin WristOx2 3150?
    • M
      Melissa TownsendSep 23, 2025
      If the %SpO indicator and the heart in the pulse strength indicator do not display on your Nonin Medical Equipment, use nVISION software to configure the device to full or partial display mode, because the device has been set to Memory Volume (MVI) display mode.
  • N
    nlevyAug 23, 2025
    What does error Er 03 or Er 05 mean on Nonin WristOx2 3150 Medical Equipment?
    • M
      Mark JonesAug 25, 2025
      If 'Er 03' or 'Er 05' is displayed on the LCD of your Nonin Medical Equipment, it indicates a device failure. The device memory is intact, but the device may have lost the most recent session or stored data.
  • T
    Tammy LarsonAug 25, 2025
    What does error Er 02 or Er 04 mean on Nonin WristOx2 3150 Medical Equipment?
    • A
      Ashley WilsonAug 26, 2025
      If 'Er 02' or 'Er 04' is displayed on the LCD of your Nonin Medical Equipment, it indicates a device memory failure.
  • F
    Francisco MatthewsAug 27, 2025
    What does error Er 03 or 05 mean on Nonin Medical Equipment?
    • C
      Cynthia AndrewsAug 27, 2025
      If error code Er 03 or Er 05 displays on the LCD of your Nonin Medical Equipment, the device has experienced a device failure. If the error code continues, contact Nonin Technical Service.
  • B
    bhollowayAug 27, 2025
    What does error Er 02 or 04 mean on Nonin Medical Equipment?
    • W
      williamshawAug 27, 2025
      If Er 02 or Er 04 displays on the LCD of your Nonin Medical Equipment, the device has experienced a memory failure and you should contact Nonin Technical Service.
  • R
    Robert FloydSep 16, 2025
    Why %SpO and heart indicator do not display on Nonin WristOx2 3150?
    • T
      Thomas ScottSep 16, 2025
      If the %SpO indicator and the heart in the pulse strength indicator do not display on your Nonin Medical Equipment, use nVISION software to configure the device to Full Display Mode, as the device may have been set to Memory Volume (MVI) Display Mode.
  • V
    vargasrobertoSep 13, 2025
    What to do if Nonin WristOx2 3150 shows poor pulse signal?
    • C
      Christine DavisSep 13, 2025
      If the poor pulse signal indicator displays and the pulse strength indicator shows two bars or less on your Nonin Medical Equipment, reposition or replace the sensor, or place the sensor on a different finger. Warm the sensor application site if hands are cold.
  • J
    Jason WashingtonSep 14, 2025
    Why Nonin WristOx2 3150 does not record in Programmed Activation Mode?
    • C
      Cindy WagnerSep 14, 2025
      If your Nonin Medical Equipment device does not record in Programmed Activation Mode, use nVISION software to program the correct start and stop times. Also, use nVISION software to reset the clock if the clock settings are lost after replacing the batteries.
  • S
    Sarah AdamsSep 5, 2025
    Why %SpO and pulse rate do not display on Nonin WristOx2 3150?
    • K
      Kelly MartinSep 5, 2025
      If the %SpO and pulse rate are not displaying on your Nonin Medical Equipment, the device may be set in Partial Display mode. Use the nVISION software to select Full Display mode and reconnect the sensor.
  • V
    Victoria ColemanSep 3, 2025
    Why does my Nonin WristOx2 3150 show a poor pulse signal?
    • T
      Tina EstradaSep 3, 2025
      If your Nonin Medical Equipment is displaying a poor pulse signal, it may be due to excessive patient motion. Try to reduce the patient's movement.

Summary

Warnings

Cautions

Displays, Controls, and Indicators

Modes

Installing Batteries

Attaching the Wristband

Attaching the Sensor

Patient Application

Verifying Operation

Startup Sequence and Self-Test

Describes the startup sequence and self-test of the device.

Error Codes

Troubleshooting

Care and Maintenance

nVISION Software

nVISION Settings

Details settings that can be programmed using nVISION software.

Accessing nVISION Settings

Step-by-step guide to access and change device settings via nVISION.

Cable Connection

USB Driver Installation

Windows 7 Driver Installation

Steps to install the USB driver for Windows 7.

Windows 8 Driver Installation

Steps to install the USB driver for Windows 8.

Windows 10 Driver Installation

Steps to install the USB driver for Windows 10.

Bluetooth Connection

Parts and Accessories

Sensors

Lists available reusable pulse oximeter sensors.

Service, Support, and Warranty

Service and Support

Information on how to contact Nonin for service and support.

Equipment Response Time

Specifications

Oximeter Specifications

Details specific performance parameters of the oximeter.

Nonin WristOx2 3150 Specifications

General IconGeneral
BrandNonin
ModelWristOx2 3150
CategoryMedical Equipment
LanguageEnglish