NONIN OXIMETER TROUBLESHOOTING
Nonin 8500, 8500M, 8500MA
1) Unit will not turn on:
a) Install six new batteries, ie Duracell, Energizer ect (avoid off brand low quality batteries)
b) Verify correct polarity of the batteries (FIGURE 2)
c) Verify battery door has metal contact
d) Disconnect any accessories attached to the device
e) Verify battery contacts, inspect for corrosion / clean contacts
2) Unit will not retain memory:
a) Verify Model # of the device, must have “M” in the model # to have memory capability
(8500M, 8500MA, 920M)
b) Has the device lost battery power for more than 2 minutes, unit will clear memory after 2
minutes with no battery power.
c) Has more than 2 minutes of data been recorded.
d) Unit must have 6 new good batteries (no flashing displays), good sensor reading 2
minutes or more.
3) Unit will not display readings:
a) Verify that only Nonin accessories are being used
b) Verify integrity of all batteries, is display blinking
c) Try another sensor
d) If 8500I is being used, remove and try again using only the sensor.
e) Verify that tester \ patient is not wearing fingernail polish
4) Unit is displaying erratic readings
a) Follow steps in item #2 above
5) Numeric displays blink at one-second intervals (Low Battery Indication)
a) Install six new batteries
b) Verify correct installation (polarity) of ALL batteries (FIGURE 2)
Nonin 8500, 8500M, 8500MA Battery placement configurations
FIGURE 2
Battery installation (6) Left to Right - #1, #3, #5
#2, #4, #6

Other manuals for Nonin 8500

Questions and Answers

  • R
    Renee GreenSep 23, 2025
    What to do if segments of SpO2 are missing on Nonin Medical Equipment?
    • C
      Cindy WilsonSep 23, 2025
      If segments of the SpO2 or display are missing, it indicates defective LED displays. The displayed values may not be reliable, so discontinue use of the device.
  • J
    Jessica GonzalesAug 25, 2025
    What to do if the displayed pulse rate on Nonin Medical Equipment doesn't correlate with the ECG monitor?
    • S
      Sandra LeblancAug 25, 2025
      If the displayed pulse rate does not correlate to the pulse rate displayed on the ECG monitor, it could be due to excessive motion at the sensor site (eliminate or reduce the cause of the motion or reposition the sensor), the patient may have an arrhythmia (assess the patient), a non-compatible sensor is being used (replace the sensor with a Nonin- branded PureLight sensor), or the ECG monitor may not be functioning properly (assess the patient).
  • M
    Michelle BoyerAug 25, 2025
    What does it mean if the Pulse Quality LED is blinking amber with each pulse on my Nonin Medical Equipment?
    • C
      Connor MartinAug 26, 2025
      If the Pulse Quality LED is blinking amber with each pulse, it indicates that the quality of the pulse signal at the sensor site is inadequate. Assess the patient and try repositioning the sensor or selecting an alternate sensor site.
  • E
    Eric BrownAug 26, 2025
    What to do if the Pulse Quality display is blinking red and the SpO and/or Pulse Rate displays are dashes on Nonin Medical Equipment?
    • K
      Kyle MillerAug 27, 2025
      If the Pulse Quality display is blinking red and the SpO and/or Pulse Rate displays are dashes, it means there is an inadequate signal at the sensor site (assess the patient and reposition the sensor or select an alternate sensor site), inadequate pulse signal due to excessive motion (reduce patient motion and reposition or relocate the sensor), or sensor failure (replace the sensor).
  • S
    Stephanie CastilloSep 23, 2025
    What to do if Nonin Medical Equipment printer is not printing after changing Prn mode?
    • B
      brucebrittanySep 23, 2025
      If the printer is not printing after changing the Prn mode, reset the pulse oximeter by turning it off and then on.
  • H
    hsmithSep 13, 2025
    How to fix Nonin Medical Equipment 8500 that will not power up?
    • E
      Elizabeth WilsonSep 13, 2025
      If your Nonin 8500 device isn't powering up, the issue might be due to: * Depleted batteries: Replace all six batteries. * Incorrect battery installation: Verify the battery sequence. * Faulty battery door: Replace the battery door if the metal contact is missing.
  • T
    Trevor GravesSep 10, 2025
    What does it mean if the Pulse Quality LED is blinking amber with each pulse on Nonin Medical Equipment?
    • M
      Meghan CookSep 10, 2025
      If the Pulse Quality LED is blinking amber with each pulse, this indicates that the quality of the pulse signal at the sensor site is inadequate. Assess the patient. Reposition the sensor or select an alternate sensor site.
  • S
    Steven RuizSep 12, 2025
    How to deal with Degradation of device performance of Nonin 8500 Medical Equipment?
    • S
      Sarah JacksonSep 12, 2025
      Degradation of device performance can be caused by electromagnetic interference (EMI). Remove the device from the EMI environment.
  • J
    Johnathan MurphySep 4, 2025
    What to do if numeric displays are blinking at once per second on Nonin 8500?
    • L
      Linda WallaceSep 4, 2025
      If the numeric displays are blinking once per second, this may indicate low batteries (replace all 6 batteries) or incorrect battery installation (verify battery orientation).
  • R
    Ronald HunterSep 1, 2025
    What to do if numeric display segments are missing on Nonin 8500?
    • J
      Judy RichardsonSep 1, 2025
      If numeric display segments are missing, it indicates defective LEDs. Discontinue use of the device.

Summary

Nonin Oximeter Troubleshooting

Unit Will Not Turn On

Troubleshoot a Nonin oximeter that fails to power on, checking battery installation, polarity, and contacts.

Unit Will Not Retain Memory

Diagnose memory retention issues in Nonin oximeters, verifying model number, battery power, and data recording.

Unit Will Not Display Readings

Address Nonin oximeter display problems by checking accessories, battery integrity, sensor type, and fingernail polish.

Erratic Readings Troubleshooting

Resolve erratic readings on Nonin oximeters by following memory retention troubleshooting steps.

Low Battery Indication

Understand and resolve the low battery blinking indicator by replacing and correctly installing batteries.

Nonin 8500 Series Battery Placement

Battery Installation Configurations

Guide to correct battery installation and polarity for Nonin 8500, 8500M, 8500MA models using Figure 2.

Nonin 8500 Specifications

Operating Environment IconOperating Environment
operating temperature range-20 to +40 °C
storage temperature range-40 to +70 °C
operating humidity range10 to 90% noncondensing
storage humidity range10 to 95% noncondensing
operating altitudeUp to 4, 000 meters
hyperbaric pressureUp to 4 atmospheres
Power IconPower
battery typeSix 1.5V AA-size alkaline batteries
battery life at maximum display brightness100 hours
battery life at normal display brightness160 hours
Display and Indicators IconDisplay and Indicators
oxygen saturation display range0 to 100% SpO2
pulse rate display range18 to 321 BPM
pulse quality indicatorLED, tricolor
numeric displays3-digit 7-segment LEDs, red
Measurement Specifications IconMeasurement Specifications
red wavelength660 nanometers
red output power0.8 mW max. avg.
infrared wavelength910 nanometers
infrared output power1.2 mW max. avg.
Weight and Dimensions IconWeight and Dimensions
width8 cm
height15 cm
depth2.5 cm
weight280 g