13
Ketone Interferences:
The Nova StatStrip Xpress-i Glucose and ß-Ketone Hospital
Meter exhibits no interference for ß-Ketone from the
following substances up to the following concentration levels:
Substances Concentration
mg/dL(mmol/L)
Acetaminophen 20.0 (1.32)
Acetone 10 (1.72)
Aceteactate 10 (0.93)
Ascorbic Acid 20 (1.14)
Bilirubin 10 (0.18)
Captopril 10 (0.46)
Cholesterol 500 (12.9)
Creatinine 6 (0.53)
Dopamine 10 (0.53)
Ephedrine 0.9 (0.035)
Glucose 900 (50.0)
Introduction
Substances Concentration
mg/dL(mmol/L)
Ibuprofen 48 (2.33)
L-Dopa 10 (0.51)
Methyl-Dopa 1 (0.042)
N-Acetyl-L-Cysteine 10 (0.61)
Tetracycline 30 (0.62)
Tolazamide 15 (0.48)
Tolbutamide 45 (1.67)
Triglycerides 75 (0.88)
Salicylate 30 (1.87)
Uric Acid 20 (1.05)