Important Information for Control Solution
• UseonlyNovaMaxControlSolutionforthetest.
• Checktheexpirationdateonthecontrolsolutionvial.Donot
use control solution past the expiration date or you may get
inaccurate results.
• Storeonlyfor3monthsafterrstopening.Whenyouopen
a new vial of control solution, count forward 3 months and
write that date on the label of the control solution vial. Dis-
card any remaining solution after the date you have written
on the vial.
• Storethecontrolsolutiontightlyclosedatroomtemperature
below86°F(30°C).Donotrefrigerateorfreeze.
• Shakethecontrolsolutionwellbeforeusing.
Caution: The Nova Max Control Solution range printed on the test
strip vial is for control solution only. It is used to test the perfor-
mance of the monitor and test strip. It is not a recommended range
for your blood glucose level.
If your control solution test results continue to fall outside the
range printed on the test strip vial:
• TheNovaMaxBloodGlucoseMonitormaynotbeworking
properly.
• Donotusethemonitortotestyourblood.
• CallCustomerServiceat1-800-681-7390.