EasyManuals Logo

Nova Max User Manual

Nova Max
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
14
Testing a Quality Control Solution
1. Insert a test strip into the monitor. If monitor was o, the
screen displays all segments for 2 seconds then the blinking
blood drop symbol appears.
NOTE: If the strip is removed before you start the test, the screen
goes blank.
2. Press the left/right
buttons to indicate this
sample is a control. (CTL is shown on the display.)
NOTE: It is important to select control solution test so the test result
does not appear to be one of your blood glucose test results.
NOTE: If a test is not performed within 2 minutes from the insertion
of the test strip, the screen goes blank. To per-
form a test, take out then replace the test strip
starting from Step 1.
3. Shake the control solution vial. Discard
adropbeforeuse.Squeezeadropof
control solution onto a clean, hard, dry
surface, i.e., control cap.
4. Pick up the monitor with test strip inserted
and touch the test strip to the control solu-
tion drop.
Running Control Solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nova Max and is the answer not in the manual?

Nova Max Specifications

General IconGeneral
BrandNova
ModelMax
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals