NOxBOX
i
®
& NOxMixer Operating Instructions
1
Set- Up
Before you use the NOxBOX Ltd NOxBOX
i
intelligent Nitric Oxide delivery and monitoring system it is important to follow these steps:
Ensure the device does not require maintenance (monthly) and that the maintenance is not due during the expected duration of the therapy such as:
NO sensor high calibration
NO
2
sensor high calibration
Pump calibration
Vent calibration
Checklist
Ensure the device is clean and free from damage
Ensure the water trap is empty and has been cleaned from previous use
Ensure all single use items from previous patient have been discarded
Equipment you will need for set-up:
A NOxBOX
i
system test kit (NOXBOX-I TEST)
A NOXKIT (inspiratory limb size dependant)
Two cylinders of delivery gas
Additional items
NOxAIR Environmental monitors
EMER kit- Emergency backup
Oxygen hose for NOxMixer
A 22mm one-way valve for HFO ventilator circuits (FXS555)
NOTE: If you do not have the parts listed above or the device has not been maintained the NOxBOX
i
may not be suitable for use and you may be unable to
set-up the device. Failure to follow these procedures may result in harm to to the operator or the patient.
NOxBOX Ltd, Station Road, Harrietsham, Maidstone, Kent, ME17 1JA
+44 (0)1622 854868 +44 (0)1622 854860 ask@noxboxltd.com www.noxboxltd.com
Issue 7.0 (Software V.17.1) March 2016. Part No: LAB637
© NOxBOX Ltd
NOxBOX Limited reserves the right to change or update this literature without prior notice. Registered ofce: England and Wales. Registered No: 9563860
Question and Answer IconNeed help?

Do you have a question about the NoxBox NOxBOXi and is the answer not in the manual?

NoxBox NOxBOXi Specifications

General IconGeneral
BrandNoxBox
ModelNOxBOXi
CategoryMedical Equipment
LanguageEnglish

Summary

Alarms Troubleshooting Guide

Notifications at Switch On

High Priority Alarms During Therapy

NO Low Alarm

Details the 'NO Low' alarm, its possible causes, and recommended troubleshooting steps.

NO High Alarm

Details the 'NO High' alarm, its possible causes, and recommended troubleshooting steps.

NO₂ High Alarm

Details the 'NO₂ High' alarm, its possible causes, and recommended troubleshooting steps.

O₂ Low Alarm

Details the 'O₂ Low' alarm, its possible causes, and recommended troubleshooting steps.

Troubleshooting Common Issues

Water Trap Full Alarm

Describes the 'Water Trap Full' alarm, its consequences, and how to empty the trap.

Sample Line Block Issue

Explains the 'Sample Line Block' issue, its causes, and recommended actions.

Battery Critical Notification

Details the 'Battery Critical' notification, system power warning, and recommended steps.

Troubleshooting Critical Issues

Gas Supply and Vent Idle Issues

Covers cylinder supply critical alarms and vent idle conditions.

Critical Delivery Faults and Touch Screen Problems

Addresses critical delivery faults and issues with touch screen unresponsiveness.

Medium Priority Alarms During Therapy

Zero Calibration Requirement

Explains the mandatory daily sensor zero calibration process and its importance.

General Troubleshooting

NOxMixer Specific Troubleshooting