14
Optional
Wipe the surface with a cloth moistened with disinfectant or
ethanol with a concentration of 60-90 %.
When using disinfectant, follow the instructions given by the
manufacturer of the disinfectant.
Use a state-sanctioned disinfectant with proven
bactericidal, fungicidal, and virucidal properties.
4 Shake the coupling body to drain water and wipe off
moisture on the surface with a dry cloth (Fig. 22).
Fig. 22
The following disinfectants can be used in the United States
and Canada:
CaviCide
®
, CaviWipes
®
(manufactured by Metrex)
NOTICE
Wiping is not an alternative method of cleaning.
Please do not skip the required cleaning steps which
are previously displayed.
After cleaning, if moisture is present, wipe it off with a dry
cloth or blow it off with compressed air (≤0.35 MPa) until
there is no moisture in the interior and exterior of the
product.
When blowing off with compressed air, cover the product
with a cloth to prevent the scattering of water.

Other manuals for NSK PTL-CL-LED