15
EN FR
Coupling wrench(PTL Type ZA01040100,
KCL-LED Type 20000953 )
1 Using a soft-bristled toothbrush, clean the entire
external surface of the coupling wrench under running
water for at least 15 seconds (Fig. 23).
(Water condition: ≤38 °C, ≥3.5 L/min; water should be of
the same quality as drinking water)
Fig. 23
2 Rinse with running water.
3 Under appropriate lighting (natural office lighting, 500 lx
or higher), inspect the coupling wrench for blood or
other debris. If any visible debris remains, repeat the
process until the handpiece is visually clean.
4 Wipe off moisture on the surface with a dry cloth.
5 Wipe the surface with a cloth moistened with
disinfectant or ethanol with a concentration of 60-90 %.
When using disinfectant, follow the instructions given by
the manufacturer of the disinfectant.
Use a state-sanctioned disinfectant with proven
bactericidal, fungicidal, and virucidal properties.
The following disinfectants can be used in the United
States and Canada:
CaviCide
®
, CaviWipes
®
(manufactured by Metrex)