5–3 Packing, Sterilizing, and
Drying
1 Use an FDA-approved sterilization pouch that conforms
to ISO 11607-1, and seal the pouch. Maximum
temperature for NSK instruments is 135 °C (275 °F)
(Fig. 24).
ISO 11607-1
Fig. 24
2 Perform steam sterilization with the following conditions.
NOTICE
Wiping is not an alternative method of cleaning.
Please do not skip the required cleaning steps which
are previously displayed.
After cleaning, if moisture is present, wipe it off with a dry
cloth or blow it off with compressed air (≤0.35 MPa) until
there is no moisture in the interior and exterior of the
product.
When blowing off with compressed air, cover the product
with a cloth to prevent the scattering of water.