Type Gravity
Displacement
Pre-Vacuum (Dynamic
Air Removal)
Temperature 132 °C (270 °F) 132 °C (270 °F)
Full Cycle Time 15 min 4 min
Drying Time 30 min 30 min
CAUTION
Use an FDA-approved steam sterilizer to perform
sterilization.
Follow local rules, regulations, and guidelines
regarding the reprocessing of devices.
Do not touch the product immediately after steam
sterilization as it will be very hot and must remain in
a sterile condition.
Do not heat or cool the product too quickly. Rapid
change in temperature could cause damage to the
product.
CAUTION
Place each product in an individual sterilization
pouch for steam sterilization. This is to prevent
possible discoloration and damage to the product
due to chemical residue from other instruments.
Clean the coupling prior to sterilization.
Be sure to use sterilizers that can perform
sterilization up to 135 °C. In some sterilizers, the
chamber temperature may exceed 135 °C. Do not
use these sterilizers as failure of the coupling could
occur. Contact the sterilizer manufacturer for detailed
information about cycle temperatures.
Steam sterilization is recommended for the product.
The validity of other sterilization methods (such as
plasma sterilization or EOG sterilization) is not
confirmed.