English
7
CAUTION
• Do not sterilize by ultraviolet ray. The handpiece could discolor.
• If autoclaved with other instruments stained with chemical solution, it could strip the plating and make the surface
black.
• Alcohol disinfect the handpiece cord after every treatment.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
8. Sterilization
• Autoclave sterilization is recommended.
• Autoclave sterilization required first time you use and after each patient as noted below. Take handpiece out of the
packing bag before sterilization.
• ONLY the Tip, Handpiece and Tip Wrench can be autoclaved.
Autoclave Procedure
1
)
Clean and Remove the Tip after use.
(
Refer to 4
)
2
)
Wipe dirt and debris from the products, and wipe clean with alcohol-immersed cotton swab or cloth. Do not use a wire brush.
3
)
Insert those into the Sterilization Case or an autoclave pouch. Seal the pouch.
4
)
Autoclavable up to max. 135˚C.
Ex.
)
Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5
)
Keep the products in the Sterilization Case or autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C for more than 15 minutes is recommended by ISO17664 and EN ISO17665-1.
CAUTION
7. How to Use Tip Cover S
(
Option
)
Grip the Tip Cover S and insert it to the Tip.
To remove, grip the Tip Cover S and the handpiece & pull.
(
Fig. 11
)
The Tip Cover S is not designed for use as a Tip changing tool.
Carefully insert the Tip into the Tip Cover S. Avoid injuring
the fingers.
Fig.11
Slit
Tip
Tip Cover
Tip-Handpiece Joint
Sterilization Case
The Handpiece, Tip and Tip Wrench can be sterilized together using
Sterilization Case.
1
)
Remove the Tip from the handpiece by using Tip Wrench.
2
)
Set the Tip Wrench with Tip into the Sterilization Case.
(
You can set four
Tip Wrenches and Tips at once
)
.
3
)
Remove handpiece from the Handpiece Cord, and clean.
(
Refer to 3
)
4
)
Set the handpiece into the Sterilization Case.
5
)
Autoclavable up to max. 135˚C.
ex.
)
Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
6
)
Keep the products in the Sterilization Case or autoclave pouch to keep it clean until you use it.
Fig.12
This handpiece can be cleaned and disinfected with a Thermo-Disinfector.