NSK Varios 170 - Troubleshooting

NSK Varios 170
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
9. Troubleshooting
When trouble is found, please check the followings prior to consulting your dealer.
Problem Item to Check Cause Solution
No / Poor
vibration.
The Tip does not
generate vibration, in
spite of depressing
the Foot Control.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks.
Worn Tip. Replace the Tip.
Power has not been correctly
adjusted for the Tip.
Adjust the power on the Power Guide or
Tip case label. Do not exceed.
Failure of vibrator in the handpiece. Contact dealer.*
The Tip is bent
or broken.
Power has not been properly
adjusted for the Tip.
Adjust the power level the Power Guide or
Tip case label. Do not exceed.
The Tip is flying
away.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks.
Noise from the
handpiece.
Power has not been properly
adjusted for the Tip.
Adjust the power level on the Power Guide
or Tip case label. Do not exceed.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks.
Failure of vibration in the handpiece
or the Module.
Contact dealer.*
The handpiece
is overheating.
Power has not been properly
adjusted for the Tip.
Adjust the power level on the Power Guide
or Tip case label. Do not exceed.
The Tip is not tightened firmly. Tighten the Tip until the Tip Wrench clicks.
Failure of vibration in the handpiece
or the Module.
Contact dealer.*
No / Weak water.
The water does
not reach to the
handpiece.
Check the water circuitry and supply to the
Unit Chair.
Remove the
handpiece from the
handpiece cord. Water
does not comes out or
weak from it.
The water adjustment dial is
closed.
Turn the water adjustment dial and set it to
an appropriate water colume.
The water filter is clogged.
Contact dealer.*
No / Weak water form
the handpiece.
There may be exogenous material
in the hole.
Air it with the appliances such as
syringes from the rear of the handpiece.
If it is still unsolved, contact dealer.
Water leakage.
Water is leaking from
the joint between the
handpiece and the
cord.
O-Ring at the handpiece cord is
worn or damaged.
Replace with new O-Ring (Refer to 6 6-2
Changing O-Ring).
Handpiece
LED does not
illuminate.
(Varios 170 LUX)
Tip oscillate, but
Handpiece LED turns
on and off.
The handpiece is not connected into
the Handpiece Cord correctly.
Firmly insert the handpiece into the
Handpiece Cord inmost.
Tip oscillate, but
Handpiece LED does
not turn on.
Disconnection in the Handpiece
Cord, or failure in Module.
Contact dealer.*
* Repairs cannot be made by the customer.