Ntron OxyTX
Operation and Installation Manual
Revised:10/2013 ECN00008
Part No.:01-507
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co Meath, IRELAND
Tel: + 353 (46)9071333 • Fax: + 353 (46)9071331 •
Email: info@ntron.com • Website: www.ntron.com/ntron
Ntron Ltd, Dallam Court, Dallam Lane, Warrington WA2 7LT, UNIITED KINGDOM
Tel: + 44 (161) 930 8690 • Fax: + 44 (161) 930 8691
TECHNICAL SPECIFICATIONS
OxyTX:
Transmitter Enclosure: ABS with Stainless Steel Identification label (IP66)
Weight: 800g
Power Requirements: 7 - 30 VDC 2 wire loop power
Output: 4 - 20 mA
Range: 0 - 25 mA
Display: 4 digit LCD display
Resolution: 0.1% Oxygen
Operating Humidity: 0 - 95% RH, non-condensing
Operating Temperature: -20°C to 45°C
Approvals: ATEX, Ex II 2 G, EEx ia IIC T4
Oxygen Sensor - Sample:
Flow: 100 ml/min minimum
Pressure: 4 PSIG maximum
Temperature: -5°C to 50°C
Oxygen Sensor:
Type: Electrochemical fuel cell
Sensor life: Typically 12 months
Operating Humidity: 15-90%
0-99% intermittent
Calibration Adjustments: Zero and span, located on front panel.
Certification: ATEX, Ex II 1 G, EEx ia IIC T6
P