EasyManua.ls Logo

Ntron OxyTX - Zero Sensor Calibration (Menu Mode C:1)

Default Icon
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ntron OxyTX
Operation and Installation Manual
Revised:10/2013 ECN00008
Part No.:01-507
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co Meath, IRELAND
Tel: + 353 (46)9071333 • Fax: + 353 (46)9071331 •
Email: info@ntron.com • Website: www.ntron.com/ntron
Ntron Ltd, Dallam Court, Dallam Lane, Warrington WA2 7LT, UNIITED KINGDOM
Tel: + 44 (161) 930 8690 • Fax: + 44 (161) 930 8691
The ‘FGD3’ menu system is as follows: -
Cal Number Function
1 Zero Sensor
2 Sensor Span
3 Select FSD
4 Setting 4mA Level
5 Setting 20mA Level
6 Display Mode
7 Display Firmware Version
8 Restore
9 View Engineer/Diagnostics Data
12 Cross Reference
19 Positive Zero Suppression
20 Negative Zero Suppression
30 Password
Zero Sensor (Menu Mode C:1):
This calibration feature allows the instrument to determine the sensor output under
zero gas conditions.
Apply zero gas (zero grade/UHP Nitrogen) to the sensor inlet and allow enough
time for the sensor to respond and all the gas to be purged (typically 2 minutes
minimum dependent upon flow rate).
Select menu mode C: 1 and press ENTER.
Press ENTER to perform the ZERO calibration. Pressing MENU instead of
ENTER aborts the calibration (the ZERO factor will still be displayed on exit).
Press MENU – the display will show the ZERO factor for the instrument before
returning to its standard mode of operation.
The ZERO factor should be recorded on any calibration certificates completed.
P
AGE 9 OF 19

Related product manuals