OM0135
13
7.3 Shelf Plenums & Water Pan Installation
Before installation of the shelves, and water pan, NuAire recommends to decontaminate all surfaces within
the interior chamber, glass door, and outer door with gasket. They can be wiped down with a disinfectant
of 70 percent alcohol or similar non-corrosive antimicrobial agent. Use only disinfectants that are
compatible with the vinyl gasket and the powder coat paint on the outer door. If the user desires, the top
and side plenums can be removed to disinfect the chamber surfaces that they cover. Absolutely no
chlorinated or halogen materials are to be used in the chamber.
Provided with each Incubator, are four shelves. The shelves are easily installed by attaching the shelf
guide supports to the sidewall plenums. If the shelf does not bind or disengage from the horizontal bracket
it is installed correctly. Additional shelves and shelf guides are available. The water pan shall be placed
directly on the bottom of the chamber.
7.4 Electrical
The electrical supply circuit to the Incubator must conform to all national and local electrical codes.
Consult the serial-data plate, located at the front of the right side of the Incubator, for voltage, cycle, phase,
and ampere requirements before making connection. Plug the power cord securely into a grounded power
source. VOLTAGE SHOULD NOT VARY MORE THAN 5% FROM SERIAL PLATE RATINGS.
Have a qualified technician check with the power source with a properly rated volt meter if needed. A
separate branch circuit is recommended to prevent possible loss of product due to overloading or failure of
other equipment on the same circuit. A SURGE PROTECTOR IS STRONGLY RECOMMENDED to
avoid power-related faults.