21
SOURCE CALIBRATION & CORRESPONDING VOLTAGE, CURRENT LOOP VALUES :
CPS
Expected
Dose rate
Observed Dose rate
18
90
180
360
900
1800
1 mR/hr
5 mR/hr
10 mR/hr
20 mR/hr
50 mR/hr
100 mR/hr
Over Load Test
Passed Status
Yes
No
(a) 10 Times Highest Range
(b) 100 Times Highest Range
4.MECHANICAL INSPECTION (HARDWARE FIXTURES)
No
Description
Observation
1.
Inside the unit
Check proper fixing of PCB and all
connectors.
OK/NOT OK
2.
Labels of main unit
Check for labels of front and side panels of
each connector.
OK/NOT OK
3.
Labels of GM probe
Check for the labels of GM probe.
OK/NOT OK
4.
Dropping test
Dropping test from height of 1meter 6times
on thermocol and recheck for proper
operation
OK/NOT OK
5.
Packing quality
Check for proper packing of the unit with
thermocol sheet.
OK/NOT OK
6.
Clamps
Check the fixed clamps are supplied with
unit for wall mountable.
OK/NOT OK
7.
Manual and spares
Check for proper user manual and spares
supplied with unit.
OK/NOT OK
TESTED BY: APPROVED BY:
DATE: DATE: