Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic environment – guidance
Table 2 (Part 1)
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m Portable and mobile RF communications
equipment should be used no closer to any part
of FeverSmart™ Temperature Monitor, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance.
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
(Continue on next page)
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLE
22