Guidance and manufacturers declaration – electromagnetic immunity
Immunity Test IEC 60601 Test Level
Table 1
Compliance Level
Electromagnetic environment – guidance
21
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLE
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-6
3 A/m 0.3 A/m
The power frequency magnetic field should be
measured in the intended installation location
to assure that it is sufficiently low.