Oakton PC 700 - 3.7 ACAL (Auto Conductivity Calibration)-Con & TDS only; 3.8 SPC (Single Point Calibration)-Con & TDS only; 3.8 CELL (Nominal Cell Constant)-Con & TDS only

Oakton PC 700
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Press .
Press or to select the desired value (15.0 to 30
Press
.0).
to confirm.
ACAL to nductivity Calibration)—Con & TDS
only9.11
3.7 (Au Co
See Sectio for more deta atic Conductivity Calibration.
ress
n 5.1 ils on Autom
. P
Press or to select “Yes” (Automatic) or “No” (Manual).
Press to confirm.
9.12 3.8 Po
See Sectio for details o
n Single Point & Multi Point Calibration.
ress
SPC (Single int Calibration)—Con & TDS only
n 5.2
. P
Press or to select “Yes” (Single) or “No” (Multi).
Press to confirm.
9.13 3.8 No al
The PC 700 includes a probe with a nominal cell constant (k) of 1.0.
Use prob k parately) for improved
performa ce in extreme sample ranges. Use this setup function to
atch the
CELL ( min Cell Constant)—Con & TDS only
es with = 0.1 and 10 (sold se
n
change the cell constant if necessary. Meter default is 1.0 to m
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
Press
.
Press or to select 0.1, 1.0, or 10.0
Press to confirm.
15