- 20 -
Slope Calibration
Use the ▲ or ▼ keys to adjust the value to
match known temperature standard. .
Press F1 to accept. Then move to second
temperature point using the ▲ or ▼ keys
and repeat.
Match Calibration
Use the ▲ or ▼ keys to adjust the value to
match T1 readings and T2 readings.