Battery Charging
Adjustment
Battery charger output voltage under normal charging operation should be adjusted to produce 225.0V at TP1 relative to TP3
using R71. TP3 is not installed. Use meter probe to connect to the exposed hole at this location. The trimpot R71 turns
counterclockwise to increase, or clockwise to reduce the output voltage. Check that high charge mode is not engaged. If the
battery charger output voltage is too high in normal charge mode, this can cause the error message PRECHARGE VOLTAGE
ERROR, and prevent completion of precharging.
Miscellaneous
None