Diagnostics
Accumulate Exposure Time
Software maintains an accumulated exposure time for the current patient. This time appears on each C-Arm Control Panel
centered between kV and mA when the system is in Manual or Auto Fluoro mode. You can manually reset the exposure time, or
let it reset automatically when the system boots or when the current patient changes.
C-Arm Events Log
System software logs generator events and corresponding data to the Generator Events Log. This is valuable information that
has great troubleshooting value. When an error message appears on the C-Arm’s Control Panel Display, use RUS/RUT to
access the Generator Events Log and review the messages there. Refer to the program help file for information on RUS/RUT
program operation. Besides listing the generator event, each message includes one or two data items. Following is an
alphabetized list of all possible messages:
Message Type or Code Data Item
AD CHANNEL %s FAIL – POWER OFF –
WAIT 5 SECONDS
Error
1-Channel number
2-Scaled A/D Value
ANODE IS HOT
Error
1-Percentage of Capacity
2-Heat Units in kilo-units
ANODE IS WARM
Error
1-Percentage of Capacity
2-Heat Units in kilo-units
HLF OVERTIME Operation 1-None
2-None
FILAMENT CAL REQUIRED Error 1-None
2-None
CAMERA IRIS FAILURE Error 1-0 = open; 1 = close
2-None
PRE-CHARGE VOLTAGE ERROR – POWER Error 1-Battery Voltage (Inverter B A/D)
16
Service