Calibration
14. Highlight and click on Save Filament Calibration Data File As.
15. When the new window appears, type in XXXXXXXX.fcl (Mainframe’s Barcode Number) then click on Save. Verify
Calibration Data File was saved.
16. Click on File located in the upper left-hand corner.
17. In the new window that appeared, highlight and click on Download Filament Calibration Data To 9800.
18. Verify that the Filament Calibration Data File was downloaded to the 9800, by checking for the Download Complete in the
lower right hand corner of the RUT window.
Using UtilitySuite
1. Attach the Serial Com Port of the Laptop to the ARCNET Connection (J4) of the Workstation. On the Laptop, point and
double click on the tool Icon and go to RUT.
2. In Main Menu Window click on Connect->System.
3. In the left-hand column, click on Filament Calibration.
4. When the Filament Calibration Window appears, click on Take Duty Cycle Data.
5. Follow the prompt to complete the data collection.
Note: The Filament Calibration takes about 12 minutes to complete.
6. Verify that the Filament Calibration was completed without any failures. Release the X-Ray switch on the Mainframe or the
Foot Pedal to stop making x-rays.
7 Click on Interpolate Data. Verify Interpolation is done.
8. Click on File->Save As->Generator File.
9. When the new window appears, type in XXXXXXXX (Mainframe’s Barcode Number) then click on Save. Verify the file was
saved.
10. Click on File->Save->To System->Generator.
11. Re-boot the system.