Calibration
Monitor Setup
Note: All adjustments to the monitors and light sensor should be performed only after the monitors have reached normal
operating temperature. Usually a minimum of 1 hour is required.
Monitor Size
Note: For pre-OPIS system software, use the RUT utility CD.
Verification Using RUT
1. Connect the utility cable to the laptop and the back of the workstation.
2. Run the RUT program. From the Tools pull down menu, select Connect, then Upload and Save to save all data from the
9800 to the laptop.
3. Select the Workstation Calibration form
4. Click on the Monitor pull down button select the left monitor.
5. Click on the Test Patterns pull-down button and select Monitor/Camera Alignment. The beam alignment pattern appears on
the left monitor.
6. Verify that the pattern on the left monitor measures 10” by 10”. If the pattern does not measure 10” by 10”, the monitor needs
resizing (see Calibration procedures following).
Verification Using Utility Suite
1. Connect the utility cable to the laptop and the back of the workstation.
2. Start the Utility Suite software.
3. Select the View Pull down menu and select “RUT”.
71