Calibration
4. Select the File Pull down menu and select New then 9800.
5. Select the Connect Pull down menu and System. When asked if you want the forms updated with data retrieved from the
system, select Yes. As data is retrieved from the system, you will see the serial and barcode numbers for the system appear
on the form.
6. Select the Monitor Calibration form.
7. From the Display Test Patterns pull down, and select Beam Alignment. The beam alignment patterns will display on the left
monitor.
8. Verify that the pattern on the left monitor measures 10” by 10”. If the pattern does not measure 10” by 10”, the monitor needs
resizing (see Calibration procedures following).
Calibration Using RUT
1. If the pattern
sizing
needs adjustment use the Vertical and Horizontal sliders on the form to make the pattern measure 10 by
10.
2. If the pattern
centering
needs adjustment use the X and Y Position Sliders to center the pattern on the monitor.
3. Click on the Monitor pull down button and select the right monitor.
4. Click on the Test Patterns pull-down button and select Monitor/Camera Alignment.
5. Use the sliders on the form to size and center the right monitor.
6. From the File pull down menu select the Save Workstation Calibration File to save the new monitor calibration data.
7. From the File pull down menu select the Download Workstation Calibration data to the 9800.
8. From the Tools pull down menu select Disconnect.
9. Turn the 9800 system off, wait for about 10 seconds, then turn the system back on.
Calibration Using Utility Suite
1. To adjust monitor size, select the Enable Monitor Adjustment Turn Off Light Sensor button and use the Vertical and
Horizontal sliders on the form to make the pattern measure 10” by 10”.