Calibration
Light Sensor Verification / Calibration
Note: All adjustments to the monitors and light sensor should be performed only after the monitors have reached normal
operating temperature. Usually a minimum of 1 hour is required.
The following equipment is required to calibrate and verify the light sensor.
Light meter
Fostec light source and head with mounting bracket.
Fostec light source control PCB
Light Sensor Verification
1. Verify that the light source is producing the proper light output in the HIGH and LOW switch settings.
2. Place the light source holding bracket over the workstation light sensor.
3. Turn the 9800 system on and let it boot completely.