Calibration
9. Turn the 9800 system off, wait for about 10 seconds then turn the system back on.
10. Perform the light sensor verification test.
Light Sensor Calibration Using Rut
Note: For the following procedure, use the Utility Suite CD.
1. Verify that the Fostec Light Source is properly adjusted.
2. Place the light source holding bracket over the workstation light sensor.
3. Connect the utility cable to the laptop and the back of the workstation.
4. Start the Utility Suite software.
5. Select the View Pull down menu and select RUT.
6. Select the File Pull down menu and choose New, then 9800.
7. Select the Connect pull down menu and click on System. When asked if you want the forms updated with data retrieved
from the system, select Yes. As data is retrieved from the system, you will see the serial and barcode numbers for the
system appear on the form.
8. Select the Monitor Calibration form.
9. Click on the Calibrate Light Sensor button at the bottom of the form and follow the instructions appearing on the laptop
simulating minimum and maximum lighting conditions.
10. While simulating, the minimum and maximum lighting conditions select the Read Light Sensor button. The reading should be
as follows:
Near 100 for maximum lighting conditions.
Near 10 for minimum lighting conditions.
11. Save the changes to the system by selecting the File pull down menu and choosing Save, then To System, then Both. The
message
YOU MUST REBOOT THE SYSTEM IN ORDER FOR THE SAVED DATA TO BE APPLIED TO THE SYSTEM
will be displayed. Click on OK.