Replacement
7. Lifting it slightly, withdraw the left Battery Pack from its battery compartment. Remove the right Battery Pack in a similar
manner.
8. Inspect each replacement Battery Pack for missing components or obvious damage.
9. Connect a temporary four-foot length of non-conductive fish tape to the wires on each replacement Battery Pack.
10. Set each Battery Pack on the floor near the rear of the C-Arm and route each fish tape up through the appropriate access
hole at the rear each battery compartment.
11. While pulling on the fish tape, gently push the left Battery Pack into place in its battery compartment. Do the same thing with
the right Battery Pack.
12. Connect the red wire with the blue terminal and black wire with the blue terminal together. Stow the connected wires under
the plastic strap identified in step 4.
13. Connect the red wire’s terminal to P3B on the Capacitor/Power module.
14. Connect the black wire’s terminal to P1A on the Capacitor/Power module.
15. Connect the Workstation’s power cord to facility’s AC outlet. Allow replacement batteries to charge for 24 hours and then
check system for proper operation.