Replacement
9. Insert the replacement cable's strain relief into the Generator chassis and tighten the setscrews.
10. Install all connectors on the Generator end of the cable and inspect your work. Be especially careful that the high voltage
cables connect properly to the High Voltage Tank.
11. Carefully adjust all cable grommets and suspension devices.
12. Power up system and check for proper operation.
High Voltage Supply Regulator PCB
The High Voltage Supply Regulator PCB is located in the Electronic Card Rack, as shown in the following illustration.
Follow these steps to replace the High Voltage Supply Regulator PCB:
1. Place the Workstation power switch in the OFF position and disconnect the Workstation's AC power cord from the AC outlet.
2. Remove the Rear Cover and disconnect the steering linkage.
3. Remove four screws that secure metal cover to the Electronic Card Rack.
4. Remove the High Voltage Supply Regulator PCB from the slot located to the far left in the Electronic Card Rack.