Replacement
Use the following procedure to replace the Power/Signal Interface PCB:
1. Turn off Workstation and disconnect its AC power plug from facility’s AC outlet.
2. Remove Left Side Cover from C-Arm.
3. Remove four screws and associated hardware that attach Power/Signal Interface PCB to C-Arm chassis. Retain hardware
for reuse.
4. Disconnect and tag connectors from J1 and P2 through P11.
5. Inspect replacement Power/Signal Interface PCB for any obvious damage or missing components.
6. Attach connectors to J1 and P2 through P11 on replacement PCB.
7. Mount replacement Power/Signal Interface PCB to chassis and secure mounting hardware.
8. Connect Workstation’s AC power plug to AC outlet.
9. Check system for proper operation.
10. Install Workstation cover removed in step 2.