Michael GreenSep 8, 2025
What to do if OHAUS Scales display 'CAL E'?
- IivanmurphySep 8, 2025
If your OHAUS Scales are showing 'CAL E', it means the calibration process has failed. Ensure you carefully follow the calibration process.
What to do if OHAUS Scales display 'CAL E'?
If your OHAUS Scales are showing 'CAL E', it means the calibration process has failed. Ensure you carefully follow the calibration process.
Why OHAUS SPX6201 Scales cannot calibrate?
If your OHAUS Scales cannot calibrate, it might be because the security switch is in a locked position.
How to fix OHAUS SPX6201 Scales that will not turn on?
If your OHAUS Scales won't turn on, the issue might be due to a lack of power. Check the power supply to ensure it is properly connected and functioning.
What to do if OHAUS SPX6201 Scales show 'Err 8.3'?
If your OHAUS Scales display the error message 'Err 8.3', it means the weight on the scale exceeds its rated capacity. To resolve this, remove the excess weight from the scale.
Why OHAUS Scales show Err 8.3?
The Err 8.3 error on your OHAUS Scales indicates an overload, meaning the weight on the scale exceeds its rated capacity.
Why OHAUS SPX6201 Scales are showing poor accuracy?
Poor accuracy in OHAUS Scales can be caused by improper calibration or an unstable environment.
What does 'Err 8.6' mean on OHAUS SPX6201 Scales, and how do I fix it?
If your OHAUS Scales display 'Err 8.6', the displayed value exceeds 999999, which can happen in Totalization mode. Reset the totalization mode to resolve this.
What does 'Err 8.1' mean on OHAUS SPX6201?
If your OHAUS Scales display the error 'Err 8.1', it means the weight exceeds the power on zero range. Make sure the weight is within the allowable range.
What does 'Err 8.4' mean on OHAUS SPX6201 Scales?
If your OHAUS Scales display 'Err 8.4', this indicates an underload condition, meaning the pan is removed. Make sure the pan is properly in place.
Why OHAUS SPX6201 show Err 8.6?
The Err 8.6 error on OHAUS Scales indicates that the displayed value exceeds 999999, which is possible in Totalization mode.
Lists essential safety instructions for installation and operation.
Details on releasing the transportation lock, installing the weighing pan, and security slot.
Steps for connecting the AC adapter and battery installation.
Importance and procedure for initial calibration after installation or relocation.
Overview of the Scout control panel and button functions.
Procedure for powering the balance on and off.
Detailed steps for performing span calibration.
Default mode for basic weighing operations.
Accessing calibration options like Span and Linearity.
Common problems, causes, and remedies for balance issues.