Questions and Answers

  • V
    Victor LyonsAug 18, 2025
    How to fix OHAUS Scout SPX Scales that will not turn on?
    • J
      Jennifer JuarezAug 18, 2025
      If your OHAUS Scales won't turn on, the issue might be due to a lack of power. Check the power supply to ensure it is properly connected and functioning.
  • S
    Sabrina HarveyAug 20, 2025
    What to do if OHAUS Scales show 'Err 8.3'?
    • J
      Jennifer SmithAug 20, 2025
      If your OHAUS Scales display the error message 'Err 8.3', it means the weight on the scale exceeds its rated capacity. To resolve this, remove the excess weight from the scale.
  • K
    khowardAug 23, 2025
    Why are my OHAUS Scout SPX Scales showing poor accuracy?
    • T
      Tiffany Tyler DVMAug 23, 2025
      If your OHAUS Scales are displaying 'Poor accuracy', it could be due to improper calibration. You should recalibrate the balance. Also, an unstable environment can cause this issue, so place the balance on a stable surface.
  • A
    Angel HunterAug 26, 2025
    What does 'Err 8.4' mean on OHAUS Scout SPX Scales?
    • M
      Melissa PorterAug 26, 2025
      If your OHAUS Scales display 'Err 8.4', this indicates an underload condition, meaning the pan is removed. Make sure the pan is properly in place.
  • B
    Bryan RamosAug 27, 2025
    What to do if OHAUS Scout SPX Scales display 'CAL E'?
    • S
      Stephen ShepardAug 27, 2025
      If your OHAUS Scales are showing 'CAL E', it means the calibration process has failed. Ensure you carefully follow the calibration process.
  • K
    kenneth02Aug 30, 2025
    What does 'Err 8.1' mean on OHAUS Scout SPX Scales?
    • O
      obrientimothyAug 30, 2025
      If your OHAUS Scales display the error 'Err 8.1', it means the weight exceeds the power on zero range. Make sure the weight is within the allowable range.
  • A
    awilkersonSep 1, 2025
    What does error 'Err 8.2' mean on OHAUS Scout SPX Scales?
    • V
      Vanessa NealSep 1, 2025
      If your OHAUS Scales display 'Err 8.2', it means the weight is below the power on zero range. Adjust the weight to meet the zero range requirements.
  • A
    allenhunterSep 4, 2025
    What does 'Err 8.5' mean on OHAUS Scales?
    • M
      Melissa LopezSep 4, 2025
      If your OHAUS Scales display 'Err 8.5', it indicates that the tare is out of range. Recheck the tare settings.
  • M
    Michael ThomasSep 6, 2025
    What does 'Err 8.6' mean on OHAUS Scales, and how do I fix it?
    • J
      Jeff HarrisSep 7, 2025
      If your OHAUS Scales display 'Err 8.6', the displayed value exceeds 999999, which can happen in Totalization mode. Reset the totalization mode to resolve this.
  • D
    Dakota MunozSep 9, 2025
    What does 'ref.err' mean on OHAUS Scout SPX?
    • D
      Deborah JosephSep 9, 2025
      If your OHAUS Scales display 'ref.err', it means there is a parts counting or percentage error because the sample weight is less than 1d. Use a sample weight that is 1d or greater.

Summary

Introduction and Safety

Safety Precautions

Lists essential safety warnings for installation, connection, and servicing.

Installation Setup

Power and Calibration

Connecting Power

Instructions for connecting the AC adapter and battery installation.

Initial Calibration

Importance and procedure for initial calibration after installation or relocation.

Operation Controls

Control Panel Overview

Identifies and explains the functions of the control panel buttons.

Basic Operations

Calibration Procedure

Step-by-step guide for performing span calibration.

Application Modes

Weighing and Counting Modes

Instructions for using the default weighing and counting modes.

Check Weighing Mode

Setup Menu

Maintenance and Support

Troubleshooting Common Issues

A table listing common symptoms, causes, and remedies for issues.

OHAUS Scout SPX Series Specifications

General IconGeneral
DisplayLCD
Power SupplyAC Adapter or Battery
Operating Temperature10°C to 40°C
Relative Humidity10% – 85%, non-condensing
Units of Measurementg, kg, lb, oz
InterfaceRS232
Application ModesWeighing, Checkweighing, Percent, Counting