EasyManua.ls Logo

OHAUS Scout Series - User Manual

OHAUS Scout Series
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Scout
TM
Series Balances - SKX
Instruction Manual

Questions and Answers

  • S
    Steven MartinAug 18, 2025
    Why my OHAUS Scales cannot turn on?
    • G
      gregory09Aug 18, 2025
      If your OHAUS Scales display 'Cannot turn on', it is because there is no power to the balance.
  • J
    Jodi LopezAug 20, 2025
    Why my OHAUS Scales cannot calibrate?
    • J
      Jordan ArnoldAug 20, 2025
      If your OHAUS Scales display 'Cannot calibrate', it might be because the security switch is in a locked position.
  • P
    Pamela SpencerAug 23, 2025
    What does Err 8.1 mean on OHAUS Scales?
    • B
      Beth WolfAug 23, 2025
      If your OHAUS Scales display 'Err 8.1', it means the weight exceeds the power on zero range.
  • A
    Andrea MasonAug 25, 2025
    What does Err 8.2 mean on OHAUS Scales?
    • J
      Joseph ChanAug 26, 2025
      If your OHAUS Scales display 'Err 8.2', it indicates that the weight is below the power on zero range.
  • T
    Theodore WilsonAug 28, 2025
    What does Err 8.4 mean on OHAUS Scales?
    • R
      Ryan ChavezAug 28, 2025
      If your OHAUS Scales display 'Err 8.4', it indicates an under load condition, which typically means the pan has been removed.
  • A
    Anthony WilliamsAug 30, 2025
    Why OHAUS Scales show poor accuracy?
    • A
      Alex FordAug 30, 2025
      If your OHAUS Scales display 'Poor accuracy', it could be due to improper calibration or an unstable environment.
  • K
    Kevin BrownSep 2, 2025
    What does Err 8.5 mean on OHAUS Scout?
    • A
      Allen StevensSep 2, 2025
      If your OHAUS Scales display 'Err 8.5', it means the tare is out of range.
  • M
    Michele CantuSep 4, 2025
    What does Err 8.6 mean on OHAUS Scales?
    • A
      Angela GilbertSep 4, 2025
      If your OHAUS Scales display 'Err 8.6', it means that the displayed value exceeds 999999, which is possible in Totalization mode.
  • K
    Kelly CamposSep 7, 2025
    What does CAL E mean on OHAUS Scales?
    • E
      Elizabeth Wong MDSep 7, 2025
      If your OHAUS Scales display 'CAL E', it means there was a failure to calibrate.
  • M
    Madison BridgesSep 9, 2025
    What does ref.err mean on OHAUS Scout?
    • K
      Kristie RoblesSep 9, 2025
      If your OHAUS Scales display 'ref.err', it means there is a parts counting or percentage error because the sample weight is less than 1d. The balance shows an error and then exits parts counting or goes to [Clr.ApU].

Summary

Introduction

Safety Precautions

Lists essential safety warnings and precautions for installation and operation of the balance.

Installation

Installing Components

Instructions to identify and assemble balance components before use.

Releasing the Shipping Lock

Details how to release the transportation lock on the balance for proper setup.

Installing the Weighing Pan

Describes the correct method for installing weighing pans, differentiating between round and rectangular types.

Connecting Power

Explains how to connect the AC adapter and install batteries for power.

Initial Calibration

Outlines the necessity and procedure for initial calibration to ensure accuracy.

Operation

Controls

Identifies and explains the functions of the balance's control panel buttons.

Calibration Operation

Describes how to perform span calibration to ensure accurate weighing results.

Menu Settings

Maintenance

Troubleshooting

Lists common symptoms, their possible causes, and remedies for balance issues.

OHAUS Scout Series Specifications

General IconGeneral
DisplayLCD with backlight
Operating Temperature10°C to 40°C (50°F to 104°F)
Overload ProtectionYes
Stainless Steel PlatformYes
ReadabilityVaries by model (e.g., 0.01g, 0.1g, 1g)
Power SupplyAC adapter or 4 AA batteries (depending on model)
CommunicationRS232
Tare RangeTo capacity by subtraction
Units of Measurementg, kg, oz, lb, pcs, %
Auto-OffYes
CalibrationExternal calibration (some models may have internal calibration)

Related product manuals