R
Richard SmithAug 18, 2025
What to do if OHAUS Scales cannot turn on?
- SSara PotterAug 18, 2025
If the OHAUS scale cannot be turned on, the cause may be no power to the scale. Verify the connections and voltage.
What to do if OHAUS Scales cannot turn on?
If the OHAUS scale cannot be turned on, the cause may be no power to the scale. Verify the connections and voltage.
How to calibrate OHAUS Valor 3000 if it cannot be calibrated?
If the OHAUS scale cannot be calibrated, it could be due to an unstable environment. Move the scale to a suitable location. Another reason may be an incorrect calibration weight; use the correct calibration weight.
What to do if OHAUS Valor 3000 shows OVER?
If the OHAUS scale displays 'OVER', it means the weight on the pan exceeds its capacity. Remove some weight from the pan.
Why OHAUS Valor 3000 shows Err 8.1?
The OHAUS scale displays 'Err 8.1' due to several reasons: the pan had a load during power on (remove weight and re-zero), the zero has drifted (run RAMP in Service Menu to check Load Cell, then calibrate), or the Load Cell is damaged (replace Load Cell).
Why OHAUS Valor 3000 shows Err 8.2?
The OHAUS scale displays 'Err 8.2' due to several reasons: the pan was removed prior to power on (install pan and re-zero), the zero has shifted (run RAMP in Service Menu to check Load Cell, then calibrate), or the Load Cell is damaged (replace Load Cell).
What to do if OHAUS Scales shows poor accuracy?
If the OHAUS scale shows poor accuracy, it may be due to improper calibration (perform calibration) or an unstable environment (move the scale to a more suitable location).
What to do if OHAUS Scales displays UNDER?
If the OHAUS scale displays 'UNDER', it means the pan was removed during weighing. Re-install the pan.
What to do if OHAUS Valor 3000 Scales cannot access mode?
If the OHAUS scale cannot access a mode, it means that the mode has not been enabled. Go to the menu and enable the mode.
What to do if OHAUS Scales shows REF Err?
If the OHAUS scale displays 'REF Err', it means the reference weight is too small. Use a larger sample.
What to do if OHAUS Valor 3000 cannot access unit?
If the OHAUS scale cannot access a unit, it means that the unit has not been enabled. Go to the menu and enable the unit.
Capacity | 3000 g |
---|---|
Readability | 0.1 g |
Display | LCD |
Housing Material | ABS Plastic |
Protection | IP54 |
Linearity | ±0.2 g |
Interface | RS232 |
Power Supply | AC Adapter |
Operating Temperature | 40 °C |
Repeatability | 0.1 g |
Units | g, kg |