Zeroing an AirAware is the act of setting 
the  instrument  to read  either 000  ppm  or 
00.0  ppm  (20.9  for  oxygen)  in  a 
gas/hazard-free  environment.  The 
atmosphere  must  be  clean  and  free  from 
the target gas during the zeroing process. 
If through ventilation and other measures 
it  is  impossible  to  entirely  remove  the 
target  gas  from  your  facility,  bottled 
“zero”  gas  should  be  applied  to  the 
instrument. 
Connect  the  zero  gas  bottle,  regulator, 
and  tubing  to  the  calibration  adapter. 
Connect  adapter  to  sensor  opening  and 
open regulator. Allow zero gas to flow for 
a minimum of 2 minutes. 
007 
With the “zero” gas still flowing to the sensor, press and hold the 
 
 key for 
a  total  of  10  seconds.  During  the  first  5  seconds  the  normal  screen  is  still 
displayed. During the next 5 seconds, the text or numeric display flashes once 
per second. If the key is still held after the full 10 seconds the word “ZERO” is 
displayed. 
A  flashing  decimal  point  on  the  lower  right  corner  of  the  display  will  appear  for  3  minutes  after  start-up, 
calibration, bump, or programming. This is an indicator that the 4-20 mA signal is locked at 3 mA and that all 
alarm functions are disabled. Once this flashing indicator clears, all functions will return to normal. 
Release the key and the zeroing process begins. At the end of the zero process, 
the  word  "PASS"  or  "FAIL"  is  displayed  for  5  seconds  along  with  a  .25 
second  audible  beep, after which  time  the  unit  reverts  to  the  normal  reading 
screen. An oxygen sensor will always be zeroed to 20.9%, regardless of what 
the span gas is set to, and no span reserve is shown for oxygen during zeroing. 
If the unit displays "FAIL" try again ensuring the area is free of the target gas. 
If  the  unit  fails  a  second  time  the  sensor  may  need  to  be  replaced.  Contact 
Industrial  Scientific  Oldham  Corporation  for a  replacement  sensor; see  page 
24 for ordering information.