5 Maintenance
23
Flip the maintenance switch (ref. 1)
into the CAL (calibration) position: the
yellow light (Figure 18, ref. 2) will be
lit and the sensor will send a 2 mA
current to the control unit
(Maintenance mode).
Verify that the sensor is located in a
clean-air environment. If not, inject
synthetic air at a flow rate of 30 l/h.
Figure 19
Wait for the measurement to stabilize
(displayed on screen) and adjust the
zero by using the ZERO
potentiometer located on the sensor
pack (ref. 2).
Inject the recommended calibration
gas at a flow rate of 30 l/h.
Wait for the measurement to stabilize.
Adjust the sensitivity by using the
sensitivity potentiometer located on
the sensor pack (rep. 1).
Figure 20
Stop injecting the calibration gas.
Remove the gas injection pipe, then
wait and verify that the signal returns
to zero. Repeat procedure if it does
not.
Flip the maintenance switch (rep. 1)
into the MES (measure) position. The
yellow light will turn off (Figure 18,
rep. 2).
Calibration is complete.
Figure 21