EasyManua.ls Logo

Oldham ctx 300 - Page 24

Oldham ctx 300
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
CTX 300
User manual
2
nd
case: CTX 300 without display (except for O
2
, SC and C0
2
)
The sensor is operating.
Flip the maintenance switch (ref. 1)
into the CAL (calibration) position: the
sensor will send a 2 mA current to the
control unit (Maintenance mode).
Verify that the sensor is located in a
clean-air environment. Use the
calibration kit and follow all
recommendations.
Figure 22
Connect a voltmeter to the AF+ and
AF- terminals (caliber mV/DC).
Figure 23
Wait for the signal to stabilize and
adjust the zero by using the ZERO
potentiometer located on the sensor
pack (Figure 24, ref. 2). The output
signal should be 0 mV.
Now inject the recommended test gas
at a flow rate of 30 l/h. Use the
calibration kit and follow all
recommendations.
Wait until the signal has stabilized.
Figure 24
Read the mV value on the voltmeter
(Figure 23), with the full scale at 1600
mV. Calculate the value to be read as
a function of your test gas.

Table of Contents

Related product manuals