Use
5 Use
This chapter describes the following:
Safety information for use
Information on how to use the product
5.1 Safety notes for use
CAUTION
Risk of injury to the patient
Contact between the HF cable and the patient can cause leakage
current.
Prevent contact between the HF cable and the patient when
HF current is activated.
CAUTION
Risk of injury to the patient
The rated accessory voltage of the TURis/TCRis resectoscope
and the bipolar hand instruments is limited. There is a risk
of sparkover leading to electrical injury, thermal injury and
unintended nerve stimulation.
Use a maximum output voltage of 1000 V
p
.
For the maximum output voltage, refer to the instructions for
use of the electrosurgical generator.
CAUTION
Risk of injury to the patient
The rated accessory voltage of the monopolar resectoscope
and the monopolar hand instruments is limited. There is a
risk of sparkover leading to electrical injury, thermal injury and
unintended nerve stimulation.
Use a maximum output voltage of 2000 V
p
.
For the maximum output voltage, refer to the instructions for
use of the electrosurgical generator.