Use
Place the patient on a dry, electrically insulating surface.
Prevent any contact between different skin surfaces (arms,
legs) of the patient. Place dry gauze between the body and
arms and between the legs to prevent such contact.
Prevent any skin contact between the patient and the user.
CAUTION
Risk of injury to the patient
Laying the HF cable in loops can lead to mechanical injury.
To prevent tripping hazards, do not lay the HF cable in
loops.
Use only plastic clips, hook fasteners and loop fasteners to x
HF cables to surgical drapes.
NOTICE
Risk of damage to the product
Coiling the HF cable can damage the product.
Do not coil the HF cable into a circle of less than 10 cm
(4 inch) in diameter.
NOTICE
Electromagnetic interference
Using this cable with other medical devices can result in higher
emission levels or reduced interference immunity.
Only use compatible equipment and accessories.
Observe the information regarding electromagnetic
compatibility in the instructions for use of the compatible
electrosurgical generator.
5.2 Using the product
Automatic presettings
When using the HF cable WA00014A, the electrosurgical
generator automatically recognizes the HF-resection electrode
and automatically applies appropriate presettings.
Instrument name: <Saline>
Mode: <SalineCut> or <PlasmaCut>