36
General information for reprocessing
Olympus recommends automated cleaning
Automated cleaning methods reduce infection risks for the
reprocessing personnel. They also provide standardized and
validated procedures.
Olympus recommends steam sterilization
Olympus recommends to use steam sterilization for the product.
7.2 Validated reprocessing methods
In accordance with ISO 17664, the following table provides an
overview over the reprocessing methods that are validated for the
microbiological efcacy of the product.
++ Method validated for microbiological efcacy and veried
for material compatibility.
– Method incompatible
* Only for “Preparation for automated cleaning”
# Not tested