EasyManuals Logo

Olympus A60003C User Manual

Olympus A60003C
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
42
Cleaning and disinfection
2. Check that no residual rinse water remains on the HF cable.
8.3 Manual disinfection
Clean the product before disinfection to remove patient
debris. Otherwise, disinfection will be insufcient. Refer to the
section “Manual cleaning” on page 40 for detailed cleaning
instructions.
Manual disinfection is not possible for the monopolar HF cables
A00010A, A00011A, A00012A, A0355, A0391, A0392.
Only use automated cleaning and thermal disinfection for these
products. Refer to the section “Automated cleaning and thermal
disinfection” on page 43.
8.3.1 Validated disinfectant
Olympus has validated manual disinfection of the product with
the following disinfectant:
Cidex Activated Dialdehyde solution manufactured by
Johnson & Johnson
45 minutes immersion time at 25 °C (77 °F)
The instructions in this chapter are based on the validated
disinfectant as specied above.
If other disinfectants are used, select immersion times,
concentrations and temperatures of the solutions as instructed by
the manufacturer of the disinfectant.
8.3.2 Manually disinfecting the product
A complete manual disinfection cycle consists of 3 steps:
Disinfection
Rinsing
Drying

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus A60003C and is the answer not in the manual?

Olympus A60003C Specifications

General IconGeneral
BrandOlympus
ModelA60003C
CategoryCables and connectors
LanguageEnglish

Related product manuals