EasyManua.ls Logo

Olympus BF-Q190 - Specifications

Olympus BF-Q190
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.2 Specifications
23
BF-190 Series OPERATION MANUAL
Ch.2
Specifications
Model BF-XP190 BF-P190
Optical system Field of view 110° 110°
Direction of view 0 (Forward viewing) 0 (Forward viewing)
Depth of field 2 – 50 mm 2 – 50 mm
Insertion section Distal end outer diameter ø 3.1 mm
(Tapered part of distal
tip: ø 2.9 mm)
ø4.2mm
Distal end enlarged
1 Objective lens
2 Light guide lens
3 Instrument channel outlet
Insertion tube outer diameter ø2.8mm ø4.1mm
Insertion section working length 600 mm 600 mm
Rotation range Right: 120
Left: 120
Right: 120
Left: 120
Instrument
channel
Channel inner diameter ø1.2mm ø2.0mm
Minimum visible distance
*1
*1 Distance from the distal end of the endoscope.
1.5 mm 3 mm
Direction from which EndoTherapy
accessories enter and exit the endoscopic
image
Bending section Angulation range UP: 210
DOWN: 130
UP: 210
DOWN: 130
Total length 880 mm 880 mm
Pre-freeze function
*2
*2 For more details, refer to the instruction manual for the CV-190.
Available Available
Electronic zoom function
*2
Not available Not available
Electronic shutter function
*2
Not available Not available
Records of endoscope’s information
*2
Available Available
NBI observation
*2
Available Available
High-frequency treatment Not Compatible Compatible
Laser treatment Not Compatible Not Compatible
2
1
3
UP
DOWN
2
1
3
UP
DOWN

Table of Contents