EPOCH 6LT — Getting Started Guide EN-5
Do not drop, hit, or otherwise abuse a battery, as doing so could expose the cell contents, which
are corrosive and explosive.
Do not short-circuit the battery terminals. A short circuit could cause injury and severe damage
to a battery making it unusable.
Do not expose a battery to moisture or rain; doing so could cause an electric shock.
Only use the EPOCH 6LT or an external charger approved by Olympus to charge the batteries.
Only use batteries supplied by Olympus.
Do not store batteries that have less than 40 % remaining charge. Recharge batteries to between
40 % and 80 % capacity before storing them.
During storage, keep the battery charge between 40 % and 80 %.
Do not leave batteries in the EPOCH 6LT unit during instrument storage.
Electrical Warnings
The instrument must only be connected to a power source corresponding to the type indicated on the
rating label.
If an unauthorized power supply cord is used to power the instrument, Olympus cannot guarantee
the electrical safety of the equipment.
To avoid the risk of electric shock, do not touch the inner conductor of the LEMO connectors. Up to
400 V can be present on the inner conductor. The warning symbol between the LEMO connectors and
AC adaptor connector shown in the figure below warns of this electric shock risk.
Equipment Disposal
Before disposing of the EPOCH 6LT, check your local laws, rules, and regulations, and follow them
accordingly.
Trademarks
All brands are trademarks or registered trademarks of their respective owners and third party entities.